Manual for VMP300-based instruments
38
1) Make sure that both versions are installed on the computer (here it is V10.12 and
V10.23).
Fig. 45: Both versions on the computer.
2) Connect the instrument
with the latest software
version (here: V10.23).
The version # is listed in
the header of EC-Lab
window.
Fig. 46: connection under the latest version.
3) Open “Firmware Upgrade/Downgrade” tool
in the “Tools” menu.
Fig. 47: Firmware
Upgrade/Downgrade tool in the
“Tools” menu.
4) The window displays the pathway of the current firmware flash (here: V10.23).
NOTE:
For V10.33 and later, the flash file is located in the “Program files” directory.