July 2010 Anthos 2010 Quick Start Guide Version 1.0
BIOCHROM ANTHOS 2010 MICROPLATE READER
QUICK START GUIDE
1. To turn on the instrument:
Connect instrument to a power source using the appropriate power cord. Turn on the
instrument using the switch at the back of the instrument.
Check user’s manual for important safety information.
2. To connect the instrument to a PC:
Connect to the PC via serial port to serial port or a serial to USB port adaptor. Determine the
communication port (com) used by the instrument. In the Start menu of the PC, go to
Control Panel\System\Hardware\Device Manager\Ports.
3. To connect instrument to ADAP software:
Insert the CD supplied with the instrument into PC, install ADAP. Open ADAP. ADAP will
prompt for a user ID and password. Login using the pre-set ID and password:
sadmin/sadmin. Once logged as sadmin, specific user IDs, passwords and administrative
rights may be set.
Select Setup/Instrument in the task bar. A dialogue box will open:
Under the Instrument tab:
In Baudrate, select 9600
In COM Port, select port or Auto
Sense
In Instrument Type, select 2010 in
the drop-down menus.
To confirm that the instrument is connected with the computer, select the Read
Configuration button. The serial number of the instrument should now appear in the
Setup/Instrument dialogue box along with compatible plate types. Select File\Save to save
settings.
Question and Answer IconNeed help?

Do you have a question about the biochrom Anthos 2010 and is the answer not in the manual?

biochrom Anthos 2010 Specifications

General IconGeneral
TypeMicroplate Reader
Read ModesAbsorbance
ShakingLinear
InterfaceRS232
Light SourceHalogen lamp
Plate TypeStandard and deep well microplates

Summary

Turn On Instrument

Connect Instrument to PC

Connect Instrument to ADAP Software

Instrument Tab Configuration

Configure instrument settings like Baudrate, COM Port, and Instrument Type in the ADAP software.

Confirm Connection and Save Settings

Verify instrument connection by reading configuration and saving settings.

Measure a Plate

Quick-Read Dialogue Box Settings

Set plate format, type, measurement position, mode, and filters for reading.

Perform Plate Measurement

Place plate, start measurement, enter plate ID, and export data.