Electronic Sphygmomanometer Users manual
14
Large cuff
32 cm42cmOptional
Medium cuff
22 cm32cmWith
Small cuff
17 cm22cmOptional
Adapter
Input
100240V, 50/60Hz 0.3Amax
Output
DC 5V, 1A
Environment specifications
Temperature
Operating: 5℃~+40;
Transportation and storage: -20℃~+55.
Humidity
Operating: 25%95%non condensing;
Transportation and storage: 25%95% (non condensing).
Atmospheric pressure
Operating: 700hPa1060hPa;
Transportation and storage: 500hPa1060hPa.
12 Troubleshooting
Symptom
Possible Reason
Solution
No response or cannot
turn on the device
1. The battery power might be at
shortage or no electricity;
1 The device itself might be
damaged.
1 Please charge the device;
2 Contact customer service.
The data cannot be
display properly or
display reads an abnormal
result
1. The cuff position was not
correct or it was not properly
tightened;
2. Speaking, moving arm or body,
being angry, excited or nervous
during test.
1. Review the cuff
application instructions
and retest;
2. Retest when calm; avoid
speaking or movement
during the test.
Cannot make connection
between device and
receiver
1. Bluetooth does not be open on
receiver;
2. The version of Android in the
receiver device isnt 4.3 or 4.3+;
3. The version of Bluetooth is not
4.0;
4. Out of connection range.
1. Open Bluetooth;
2. Upgrade the Android system
to version 4.3 or 4.3+;
3. Use the receiver device with
Bluetooth 4.0;
4. Close to receiver.