17
Appendix A Guidance and Manufacturers
Declaration of EMC
Note
This product needs special precautions regarding EMC and needs to be installed and
put into service according to the EMC information provided, and this unit can be
affected by portable and mobile RF communications equipment.
Do not use a mobile phone or other devices that emit electromagnetic fields, near the
unit. This may result in incorrect operation of the unit.
Caution:
This unit has been thoroughly tested and inspected to assure proper performance and
operation.
This machine should not be used adjacent to or stacked with other equipment and that
if adjacent or stacked use is necessary, this machine should be observed to verify
normal operation in the configuration in which it will be used.
Guidance and manufactures declaration electromagnetic emission
The device is intended for use in the electromagnetic environment specified below. The customer
of the user of the device should assure that it is used in such an environment.
Emission test
Compliance
Electromagnetic environment guidance
RF emissions
CISPR 11
Group 1
The device use RF energy only for its internal
function. Therefore, its RF emissions are very
low and are not likely to cause any interference
in nearby electronic equipment.
RF emission
CISPR 11
Class B
The device is suitable for use in all
establishments, including domestic
establishments and those directly connected to
the public low-voltage power supply network that
supplies buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies