48 MP System Hardware Guide
TSD104A BLOOD PRESSURE TRANSDUCER
RX104A Replacement Element
The TSD104A is used to measure direct arterial or venous blood pressure in animals for research or teaching. It is
designed to interface with the DA100C via a 3-meter cable (supplied). The RX104A is a replacement element for
the TSD104A blood pressure transducer; it does not include the TCI connector and cable.
TSD104A Specifications
Operational pressure: -50 mmHg to +300 mmHg
Overpressure: -400 mmHg to +4,000 mmHg
Dynamic Response: 100Hz
Unbalance: 50 mmHg max
Connection ports: Male Luer (2)
Eight-hour Drift: 1mmHg after 5 minute warm-up
Isolation: <
5 μA leakage at 120 VAC/60Hz
Defibrillation: Withstands 5 discharges of 400 joules in 5 minutes across a load
Operating temperature: +15° C to +40° C
Storage Temperature: -30° C to +60° C
Combined effects of sensitivity,
linearity, and hysteresis: 1 mmHg (nominal)
Output: 5 μV/mmHg (normalized to 1V excitation)
Weight: 11.5 grams
Transducer Dimensions: 67mm long x 25mm wide
Cable length: 3 meters
Interface: DA100C
TSD104A Calibration
See DA100C Calibration options on page
45.