EasyManua.ls Logo

BIOPAC Systems, Inc. MP150 - Page 47

Default Icon
251 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
www.biopac.com 47
REFCAL REFERENCE CALIBRATOR FOR THE DA100C
The REFCAL is used to check the reference voltage of the DA100C. It connects to the DA100C and displays
the reference voltage as an analog input signal. This makes it very easy to adjust the reference voltage of the
DA100C to suit your transducer.
The REFCAL connects the VREF1 and VREF2 voltage reference outputs directly to the DA100C inputs via a
precision attenuator of value (1/50). When using the REFCAL to set the DA100C references, the DA100C
should be set to DC with a gain of 50.
The voltage output on the selected channel of the DA100C will be the voltage difference between VREF1 and
VRREF2:
V
OUT
= V
REF1
– V
REF2
CBLCAL Calibration Cable for the DA100C
Use the CBLCAL to verify the signal calibration of the DA100C. This cable (1.8m) connects between the
DA100C input and the UIM100C D/A output 0 or 1. To verify the DA100C’s frequency response and gain
settings, create a stimulus signal with AcqKnowledge and monitor the DA100C’s output. The CBLCAL
incorporates a precision 1/1000 signal attenuator.
See also: Application Note #AH102 — Biopotential Amplifier Testing using CBLCAL

Other manuals for BIOPAC Systems, Inc. MP150