Appendix C MultiPro Sensor Information
Part No. Description Range Resolution
54-49-80 LEL Combustible Gas 0 – 100% LEL 1% LEL
54-49-90 O
2
Oxygen 0 – 30% by Volume 0.1%
54-49-01 CO Carbon Monoxide 0 – 1000 PPM 1 PPM
54-49-02 H
2
S Hydrogen Sulfide 0 – 200 PPM 1 PPM
54-49-14
Duo-Tox Dual Channel CO/H
2
S
Provides substance specific readouts for CO & H
2
S
CO: 0 – 1000 PPM
H
2
S: 0 – 200 PPM
1 PPM
1 PPM
54-49-24
Duo-Tox Dual Channel CO/H
2
S
Provides substance specific readouts for CO & H
2
S
Requires firmware version 2.05 or higher
CO: 0 – 1000 PPM
H
2
S: 0 – 200 PPM
1 PPM
1 PPM
Appendix D Toxic Sensor Cross-Sensitivity
The table below provides the cross-sensitivity response of the MultiPro toxic gas sensors to common interference
gases. The values are expressed as a percentage of the primary sensitivity, or the reading of the sensor when
exposed to 100ppm of the interfering gas at 20ºC. These values are approximate. The actual values depend on the
age and condition of the sensor. Sensors should always be calibrated to the primary gas type. Cross-sensitive
gases should not be used as sensor calibration surrogates without the express written consent of Biosystems.
SENSOR CO H
2
S SO
2
NO NO
2
Cl
2
ClO
2
H
2
HCN HCl NH
3
C
2
H
4
C
2
H
2
arbon Monoxide
CO)
100 10 5 10 -15 -5 -15 50 15 3 0 75 250
Hydrogen Sulfide
H
2
S)
0.5 100 20 2 -20 -20 -60 0.2 0 0 0 n/d n/d
Appendix E: Basic Parts List
MultiPro Remote Sampling Accessories
Part No. Description
54-49-102
Continuous sample pump for MultiPro. Includes 10’ of sample tubing and sample probe
assembly. UL/CSA-approved version.
54-49-102-5
Continuous sample pump for MultiPro. Includes 10’ of sample tubing and sample probe
assembly. ATEX-approved version (European).
54-49-104 Sample draw/calibration adapter.
54-49-105 Sample draw kit. Includes adapter, squeeze bulb, 10’ of tubing and sample probe assembly.
54-05-A0403 Sample probe assembly. Does not include tubing, squeeze bulb, or sample draw adapter.
54-05-A0405
Sample probe assembly with 11.5-inch stainless-steel probe tube. Does not include tubing,
squeeze bulb, or sample draw / calibration adapter.
MultiPro Sensors
Part No. Description
54-49-80 LEL Combustible gas
54-49-90 O
2
Oxygen
54-49-01 CO Carbon monoxide
54-49-02 H
2
S Hydrogen sulfide
54-49-14 Duo-Tox Dual channel CO/H
2
S Provides substance specific readouts for CO & H
2
S
54-49-200 Multi Pro sensor exchange set. O2, LEL, Duo-Tox***
54-49-201 Multi Pro sensor exchange set. O2, LEL, CO***
54-49-202 Multi Pro sensor exchange set. O2, LEL, H2S***
***This part number requires return of identical set of expired sensors.
MultiPro Charging & Datalogging Accessories
Part No. Description
54-49-103-1 Slip-in fast charger (110 VAC) for MultiPro. Includes “wall cube” supply (UL/CSA-Version)
54-49-103-5 Slip-in fast charger (220 VAC) for MultiPro. Includes “wall cube” supply (ATEX-Version)
54-49-103-12
Slip-in fast charger (12 VDC) for MultiPro. Includes “cigarette lighter” style adapter, and
vehicle mounting hardware)
54-49-106 MultiPro Alkaline battery pack (UL/CSA-approved version)
54-49-106-5 MultiPro Alkaline battery pack (ATEX-approved version (European))
54-49-107 MultiPro Li-Ion rechargeable battery pack (UL/CSA-Approved version)
54-49-107-5 MultiPro Li-Ion rechargeable battery pack (ATEX-approved version (European))
54-26-0603 MultiPro BioTrak Software Kit. Includes BioTrak software and manual