*RST
This command resets the electronic load to the following factory-defined states:
CURR MIN RES:TRAN:BLEV MIN
CURR:MODE FIX RES:TRAN:BWID MIN
CURR:PROT:DEL 3 RES:TRAN:MODE CONT
CURR:PROT:LEV MAX SENS:AVER:COUN 8
CURR:PROT:STAT OFF SENS:AVER:STAT 1
CURR:RANG MAX SENS:FUNC:CURR DC
CURR:SLEW MAX SENS:NPLC 7
CURR:TRAN:ALEV MAX SENS:VOLT:RANG:AUTO ON
CURR:TRAN:AWID MIN TRAC:FEED TWO
CURR:TRAN:BLEV MIN TRAC:FEED:MODE NEV
CURR:TRAN:BWID MIN TRAC:POIN 2
CURR:TRAN:MODE CONT TRAN OFF
FUNC CURR TRIG:COUN 1
FUNC:MODE FIX TRIG:DEL 0
INP OFF VOLT MAX
INP:SHOR OFF VOLT:ON MIN
POW:CONF MAX VOLT:ON:LATC ON
POW:PROT:DEL 3 VOLT:RANG MAX
RES MAX VOLT:TRAN:ALEV MAX
RES:RANG MAX VOLT:TRAN:AWID MIN
RES:TRAN:ALEV MAX VOLT:TRAN:BWID MIN
RES:TRAN:AWID MIN VOLT:TRAN:MODE CONT
NOTE:
*RST does not clear any of the status registers or the error queue, and does not affect any
interface error conditions.
*RST sets the trigger system to the Idle state.
*RST clears the presently active list.
Command Syntax
*RST
Parameters
None
Related Commands
*PSC *SAV

Other manuals for BK Precision 8601

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the BK Precision 8601 and is the answer not in the manual?

BK Precision 8601 Specifications

General IconGeneral
BrandBK Precision
Model8601
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals