52
Command Syntax
[SOURce:]VOLTage:RANGe:AUTO<bool>
Parameters
0 | 1 | ON | OFF
*RST Value
1
Examples
VOLT:RANG:AUTO 1
Query Syntax
[SOURce:]VOLTage:RANGe:AUTO?
Returned Parameters
<NR1>
[SOURce:]VOLTage:ON
This command sets the voltage of sink current on.
Command Syntax
[SOURce:]VOLTage[:LEVel]:ON <NRf+>
Parameters
0 through max. rated voltage.
Unit
V (volts)
Examples
VOLT 5
Query Syntax
[SOURce:]VOLTage[:LEVel]:ON?
Returned Parameters
<NR3>
Related Commands
VOLT:LATCh
[SOURce:]VOLTage:LATCh
This command sets the action type of Von.
Command Syntax
[SOURce:]VOLTage:LATCh[:STATe] <b>
Parameters
0 | 1 | ON | OFF
*RST Value
ON
Examples
VOLT:LATC 1
Query Syntax
[SOURce:] VOLTage:LATCh[:STATe]?
Returned Parameters
0 | 1
Related Commands
VOLT:ON
[SOURce:]VOLTage:HIGH
[SOURce:]VOLTage:LOW
These commands set the high and low current limit when the load is in constant voltage mode.
Command Syntax
[SOURce:]VOLTage:HIGH <NRf+>
[SOURce:]VOLTage:LOW <NRf+>
Parameters
0 to max. rated current.
Unit
A (amps)