Examples
VOLT:HIGH 5
Query Syntax
[SOURce:]VOLTage:HIGH?
[SOURce:]VOLTage:LOW?
[SOURce:]VOLTage:TRANsient:MODE
This command selects the operating mode of the transient generator as follows in constant voltage
mode.
CONTinuous The transient generator puts out a continuous pulse stream after receipt of a trigger.
PULSe The transient generator puts out a single pulse upon receipt of a trigger.
TOGGle The transient generator toggles between two levels upon receipt of a trigger.
Command Syntax
[SOURce:]VOLTage:TRANsient:MODE <mode>
Parameters
CONTinuous | PULSe | TOGGle
*RST Value
CONTinuous
Examples
VOLT:TRAN:MODE PULS
Query Syntax
[SOURce:] VOLTage:TRANsient:MODE?
Returned Parameters
<CRD>
Related Commands
VOLT:TRAN:ALEV TRAN
[SOURce:]VOLTage:TRANsient:ALEVel
[SOURce:]VOLTage:TRANsient:BLEVel
These commands specify the transient level of the input voltage. The transient function switches
between level A and level B.
Command Syntax
[SOURce:]VOLTage:TRANsient:ALEVel <NRf+>
[SOURce:]VOLTage:TRANsient:BLEVel <NRf+>
Parameters
0 to max. rated voltage
Unit
V (volts)
Examples
VOLT:TRAN:ALEV 5 VOLT:TRAN:BLEV 0.5
Query Syntax
[SOURce:]VOLTage:TRANsient:ALEVel?
[SOURce:]VOLTage:TRANsient:BLEVel?
Returned Parameters
<NR3>
Related Commands
VOLT
[SOURce:]VOLTage:TRANsient:AWIDth
[SOURce:]VOLTage:TRANsient:BWIDth
This command specifies the transient pulse width of the input voltage.