[SOURce:]POWer:HIGH
[SOURce:]POWer:LOW
This command sets the voltage high and low limit determined when in constant power mode.
Command Syntax
[SOURce:]POWer:HIGH <NRf+>
[SOURce:]POWer:LOW <NRf+>
Parameters
0 to max. rated voltage
Unit
V (volts)
Examples
POW:HIGH 5
Query Syntax
[SOURce:]POWer:HIGH? [SOURce:]POWer:LOW?
Returned Parameters
<NR3>
[SOURce:]POWer:PROTection
This command sets the soft power protection level. If the input power exceeds the soft power
protection level for the time specified by POW:PROT:DEL, the input is turned off.
NOTE: Use POW:PROT:DEL to prevent momentary power limit conditions caused by programmed
changes from tripping the overpower protection.
Command Syntax
[SOURce:]POWer:PROTection[:LEVel] <NRf+>
Parameters
0 to max. rated power
Unit
W (power)
*RST Value
MAXimum
Examples
POW:PROT 100
Query Syntax
[SOURce:]POWer:PROTection[:LEVel]?
Returned Parameters
<NR3>
Related Commands
POW:PROT:DEL
[SOURce:]POWer:PROTection:DELay
This command specifies the time that the input power can exceed the protection level before the input
is turned off.
Command Syntax
[SOURce:]POWer:PROTection:DELay] <NRf+>

Other manuals for BK Precision 8601

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the BK Precision 8601 and is the answer not in the manual?

BK Precision 8601 Specifications

General IconGeneral
BrandBK Precision
Model8601
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals