62
[SOURce:]LIST:SLEW
This command sets the slew rate for each step. This command programs both positive and negative
going slew rates. MAXimum sets the slew to its fastest possible rate. MINimum sets the slew to its
slowest rate.
[SOURce:]LIST:SLEW <NR1>, <NRf>
See specifications for range
LIST:SLEW 1, 1.5 LIST:SLEW 2, MAX
[SOURce:]LIST:SLEW? <NR1>
<NR1> is step number
[SOURce:]LIST:WIDth
This command sets the sequence of list dwell times. Each value represents the time in seconds that the
input will remain at the particular list step point before completing the step. If times exceed 16383S,
the input remains at the present level until a trigger sequences the next point in the list. Else At the
end of the dwell time, the input automatically changes to the next point in the list.
[SOURce:]LIST:WIDth <NR1>, <NRf>
LIST:WID 1, 0.02 LIST:WID 2, 0.5
[SOURce:]LIST:WIDth? <NR1>
<NR1> is step number
[SOURce:]LIST:SAV
This command stores the present list file of the electronic load to a specified location in memory. Up to
5 files can be stored. Files saved in locations 1-5 are nonvolatile and the data will be saved when power
is removed.