Blease Test Procedure
Title: Final Test 8500 Document No: 129TP000 Sheet 1 of 17
BLEASE QUALITY DOCUMENT
BQD 083.V2
25/5/94 Checked:-________
1.0 OBJECTIVES
To final test and provide a set of results for the 8500 Anaesthesia Ventilator.
2.0 SCOPE
The final test in conjunction with the checklist and calibration record (129TC000) should be
completed whenever an 8500 Anaesthesia Ventilator is being readied for shipment from the
factory.
This may be for Demonstration or Production purposes. It should also be used on ventilators
which are going into stock, and are to be fully packed and for shipping within a few days.
WARNING - Under no circumstances should ANY part of the ventilator be opened or
removed after this test procedure has been completed and signed off. This procedure is
only valid for software versions 8.20 & 8.60 onwards.
3.0 OVERVIEW
The remaining sections of this document are:
Section 4 Test Limits Specification
Section 5 Test Procedure
Section 6 Using the Calibration Analyser
Section 4 contains the test limits for the final calibration checks. The test results noted for
section 5 on the checklist and calibration record (129TC000) must be compared against these
test limits for conformity.
Section 5 gives details of the test and calibration to be done with a list of test equipment. The
section is written so that minimum adjustment of controls is done between each test. This
means that any particular test assumes all previous tests have been set-up IN THE
SEQUENCE STATED and passed. If this is not done, invalid results may be taken.
Section 6 gives a few pointers on using the Calibration Analyser with respect to this test
procedure. A good place to start if you have not used this unit and/or have not used this
test procedure for some time.
Issue 1 2 3 4 5 6
Change Note No n/a 3724 3829 3847 3866 3962
Originator JLB JLB JLB JLB JLB JLB
Checked DB RHC RHC RHC RHC
Date Nov 00 Dec 00 June 01 July 01 August 01