BLEASE QUALITY DOCUMENT
BQD 084.V2
10/5/96
Page 3 of 3
Ref Operation\Description Value 2 Value 1 Units
5.19 PCV Test Default Settings
Monitored Peak (Value 2) delivered Peak (Value 1)
Monitored Peak (Value 2) delivered Peak (Value 1)
Return to default
PCV Test Inspired Flow
Monitored Peak (Value 2) delivered Peak (Value 1)
Monitored Peak (Value 2) delivered Peak (Value 1)
Return to default
-------------
-------------
-------------
-------------
-------------
-------------
-------------
cm H
2
0
cm H
2
0
P/F
-------------
cm H
2
O
cm H
2
O
P/F
5.20 PEEP 500ml, 12 BPM, 1:2.0
Delivered minimum pressure (value 2), monitored PEEP (value 1)
Set 3 cmH
2
0
Set 20 cmH
2
0
Set 10 cmH
2
0
PEEP set to OFF
cm H
2
O
cm H
2
O
cm H
2
O
P/F
5.21 Hardware Pressure Overload
Cm H
2
O
5.22 Fresh Gas
0
1
3
5
10
LPM
LPM
LPM
LPM
LPM
5.23 O
2
Test
100%
50%
21%
OFF
-------------
-------------
-------------
-------------
%
%
%
P/F
5.24 Absorber
Language
O.E.M
-----------
-----------
-----------
Y/N
Lang
Name
5.25 Test completion, mode switch centre off.
Check readings, sign and date all documents
Issue No: 4 Originator: JLB Checked: Date: ECN Ref: 3829