Title: Final Test 8500 Document No: 129TP000 Sheet 10 of 17
BLEASE QUALITY DOCUMENT
BQD 083.V2
25/5/94 Checked:-________
Switch the ventilator to paediatric mode and carry out compliance compensation as directed
on screen:
Set ventilatory controls to: TV = 0.5L/50ml
Freq = 10 BPM
Ratio = 1:2.5
Note bellows movement.
Set ventilatory controls to: TV = 2L/200ml
Freq = 10 BPM
Ratio = 1:2.0
Note bellows movement.
5.17 Paediatric Pressure Limit
Set ventilatory controls to: TV = 6L/200ml
Freq = 30 BPM
Ratio = 1:2.0
Set pressure limit to 20 cm H
2
O and note the delivered peak pressure after several breaths.
Note the bellows position.
Disconnect lower patient airway connector.
Note peak pressure.
Switch the ventilator to Adult mode and carry out compliance compensation as directed
on screen: