Title: Final Test 8500 Document No: 129TP000 Sheet 17 of 17
BLEASE QUALITY DOCUMENT
BQD 083.V2
25/5/94 Checked:-________
6.4 Typical Calibration Analyser Equipment Set up
The equipment is connected as shown below. All other Calibration Analyser connections
should be left open to atmosphere.
20R
Resistor
20 L
Compliance
12600010
Calibration
Analyser
Fresh gas is set for minimum required
to maintain full bellows <0.5 l/min
Lo pressure
Inlet P+
Anaesthesia Ventilator
Bellows and Absorber
Driving
Gas
Short Hose
Hi Flow
Outlet
Hi Flow
Inlet
Patient
Circuit
Fresh Gas Sensor
Flow Sensor