G2S A20 / G2S C20 User Manual V1.0
6 / 39
Sound Pressure Level
28 dB, when the device is working at the pressure of 10 hPa.
Sound Power Level
38 dB, when the device is working at the pressure of 10 hPa.
Maximum Flow
Test Pressures (hPa)
4
15
20
Measured Pressure at the
Patient Connection Port (hPa)
3
14
19
Average Flow at the Patient
Connection Port (L/min)
85
140
140
Automatic leak compensation
The pressure error ≤ ± 1 hPa
when the air leakage is no more than 70 L / min.
Pressure
Range: 0 20 hPa
Margin of Error: ± (0.4 hPa + 4%)
Tube
Length: 6 ft. (1.83 m)
Maximum Delivered Gas Temperature
43°C
The Form and the Dimensions of the Patient Connection Port
The 22 mm conical air outlet complies with ISO 5356-1.