Guidance and manufacturer’s declaration - electromagnetic immunity
The Body Clock Stimplus Pro is intended for use in the electromagnetic
environment specified below. The user of the equipment or system
should ensure that it is used in such an environment
Immunity test IEC 60601 Compliance Electromagnetic environment guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient / burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips,
short
interruptions
and voltage
variations on
power supply
input lines IEC
61000-4-11
Power frequency
(50/60Hz)
magnetic field
IEC 61000-4-8
NOTE UT is the a.c. mains voltage prior to application of the test level
Electromagnetic environmental
guidance
Floor should be wood concrete or
ceramic tile. If floors are covered with
synthetic material, the relative
humidity should be at least 30%.
n/a
n/a
Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical commercial
or hospital environment.
±6kV contact
±8kV
±2kV for
power supply
lines
±1kV line and
neutral
‹5% UT (›95%
dip in UT) for
0.5 cycle
40% UT (60%
dip in UT for 5
cycles)
70% UT (30%
dip in UT) for
25 cycles
‹5% UT (›95%
dip in UT) for
5s
3A/m
±6kV contact
±8kV
n/a
n/a
n/a
n/a