Error Codes
Code Description Resolution
Er1 Battery low Replace batteries with four Duracell
AAs.
Er2 Movement
detected
Avoid moving or talking while taking
a measurement.
Er3 Overload The device is only accurate up to its
rated capacity.
Er4 No cellular
coverage
Move to another area, closer to a
window.
Use the device at a location where
you get strong cellular signal with
your mobile phone.
Er5 Cellular
connection
error
Er6 Data
transmission
error
Er7 Internal error Replace batteries with four Duracell
AAs.
If the issue persists, contact
customer support.
Er9 Provisioning
error
Retake measurement.
If the issue persists, contact
customer support.