EasyManuals Logo

Boonton 4530 Series User Manual

Boonton 4530 Series
38 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #9 background imageLoading...
Page #9 background image
BOONTON ELECTRONICS a WIRELESS TELECOM GROUP COMPANY
8
9) CHANNEL 1 CW ACQUISITION TESTS
a. Select internal calibrator and turn on at 0dBm:
Display the Main Menu > Calibratr menu to determine which calibrator is currently selected.
If the bottom line indicates “SelectInt”, then highlight that line and press Enter to select the
internal calibrator. If the bottom line indicates “SelectExt”, then the internal calibrator is
already selected.
Set Main Menu > Calibratr > Int Level to 0.0dBm.
Set Main Menu > Calibratr > Int Signal to “On”.
b. Plug in the CW sensor cable into the 4530 Sensor 1 connector and connect a model 51075 CW
sensor to the 4530 Calibrator.
Set Main Menu > Chan 1 > MeasMode to “CW.
Set Main Menu > Chan 1 > Params > dB Offset to 0.00dB.
Set Main Menu > Chan 1 > Params > Frequency to 0.05GHz.
Set Main Menu > Chan 1 > Params > Filter to “Auto”.
Set Main Menu > Chan 1 > Params > Range to “Auto”.
c. Zero the sensor:
Press the Zero/CAL key, and press Enter.
Select “Zero Chan”. Press Enter.
The 4530 will perform a sensor fine zero adjustment operation.
Check for any error messages.
d. Perform a Fixed Calibration:
Press the Zero/CAL key.
Select “Fixed Cal”, and press Enter.
The 4530 will perform a fixed 0dBm calibration.
Check for any error messages, then press the Text key.
Channel 1 should display 0.000 dBm +/-0.002dBm.
e. Perform an Auto Calibration:
Press the Zero/CAL key.
Select “Auto Cal”. Press Enter.
The 4530 will perform an automatic sensor power sweep calibration.
Check for any error messages, then press the Text key.
Channel 1 should display 0.000 dBm +-0.002dBm.
10) CHANNEL 2 CW ACQUISITION TESTS (Model 4532 only)
a. Select internal calibrator and turn on at 0dBm:
Display the M
ain Menu > Calibratr menu to determine which calibrator is currently selected.
If the bottom line indicates “SelectInt”, then highlight that line and press Enter to select the
internal calibrator. If the bottom line indicates “SelectExt”, then the internal calibrator is
already selected.
Set Main Menu > Calibratr > Int Level to 0.0dBm.
Set Main Menu > Calibratr > Int Signal to “On”.
Scans by Artekmedia => 2009

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Boonton 4530 Series and is the answer not in the manual?

Boonton 4530 Series Specifications

General IconGeneral
BrandBoonton
Model4530 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals