Boonton 9240 Series RF Voltmeter
Remote Operation
5-21
5.5.9 OUTPut Subsystem
The OUTPut group of commands is used to control various outputs of the 9240. These outputs include the Recorder Output.
OUTPut commands for the Recorder Output include setting the DC output to the MAXimum or MINimum level or can be
forced to any level in between. Also the Recorder Output can be assigned to either channel 1 or 2.
OUTPut:RECorder:FORCe
Description: Command sets the output voltage to the argument. Query returns the output voltage previously set
by FORCe or causes error -221 “Settings conflict” if the Recorder Ouput has been assigned to a
channel.
Syntax: OUTPut:RECorder:FORCe <numeric value>
Argument: <numeric value> = 0.000 to + 10.000 V
OUTPut:RECorder:MAX
Description: Set or return the recorder output maximum, or full scale (+10.0V) voltage reference level.
Syntax: OUTPut:RECorder:MAX
Argument: none
OUTPut:RECorder:MIN
Description: Set or return the recorder output minimum, or downscale (0.0V) voltage reference level.
Syntax: OUTPut:RECorder:MIN
Argument: none
OUTPut:RECorder:SOURce
Description: Set or return the source channel for the Recorder Output.
Syntax: OUTPut:RECorder:SOURce <character data>
Argument: <character data> = CH1, CH2, ACTIVE