Vercise
TM
DBS Charging Handbook
91171709-15 Page 12 of 109
Vercise
TM
DBS Charging Handbook
en
Properly Aligning the Charger
1. Press the power button. The indicator light will come on again,
and the Charger will begin beeping repeatedly to signal that it
is searching for the Stimulator.
2. Position the Charger over the Stimulator.
The beeping will stop when the Charger is partially aligned
with the Stimulator. For optimal charging, make sure the
Charger is centered over the Stimulator (Figure 4).
Charging time may increase if the Charger is not fully
aligned with the Stimulator. To fully align the Charger over
the Stimulator, position the bottom curve of the Charger
approximately 1 cm lower than the bottom curve of the
Stimulator.
If you are having trouble properly aligning the Charger or experiencing longer than normal charging times, see the “Tips for
Properly Aligning the Charger” section of this manual.
Figure 4. Position the Charger Over the Stimulator

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Boston Scientific Vercise DB-64125 and is the answer not in the manual?

Boston Scientific Vercise DB-64125 Specifications

General IconGeneral
BrandBoston Scientific
ModelVercise DB-64125
CategoryBattery Charger
LanguageEnglish

Related product manuals