BpTRU Operator’s Manual
34
Symptoms
Possible Causes
Corrective Actions
Buzzer does not sound
at end of sixth
measurement in
Automatic mode.
BpTRU is set to BP
Silent mode.
Exit BP Silent mode by
turning off power and
then holding Stop
button down while
turning power back
on.
Buzzer does not sound
at end of single
temperature
measurement.
BpTRU is set to
Temperature Silent
mode.
Exit Temperature
Silent mode by turning
off power and then
holding Temperature
Start button down
while turning on
monitor.
Cuff does not inflate
during operation.
Cuff or tubing is not
attached properly.
Check all cuff and
tubing connectors for
tightness.
Perform a BP Zero
Calibration Check
(page 71) or BP
Reference Gauge
Calibration Check
(page 72).
Leakage in cuff or
tubing.
Check cuff and tubing
for cracks and tears
which may lead to
leakages
Perform a BP Zero
Calibration Check
(page 71) or BP
Reference Gauge
Calibration Check
(page 72).
Internal leakage.
Contact BpTRU
Customer Support.