Specifications
Model: BP 2510
Operation principle: Oscillometric method
Display: Liquid Crystal Display
Measuring range: 0 – 300 mmHg (cuff pressure)
40 – 160 / min (pulse rate)
Blood pressure 50 mmHg (minimum diastolic value)
measurement: 250 mmHg (maximum systolic value)
Laboratory accuracy: ± 3 mmHg (cuff pressure)
± 5 % of reading (pulse rate)
Clinical accuracy: According to AAMI-SP 10 with auscultatory reference:
< 5 mmHg systematic offset
< 8 mmHg standard deviation
Inflation: Automatic, adapted to blood pressure
Deflation: Electronic control valve adapted to heart frequency
Power supply: 2 batteries, type LR 03 (AAA)
(e.g. Duracell Ultra)
Operating temperature: + 10 °C to + 40 °C
Storage temperature: – 20 °C to + 60 °C
Humidity: Up to 85 % RH
Cuff: Suitable for wrist circumference of 13.5 to 20 cm
Clinical validation: According to AAMI-SP10 with auscultatory reference.
(Please note that a validation against intra-arterial measurements can lead to different
results for devices which are calibrated against auscultatory reference.)
Type BF equipment
Subject to change without notice.
This appliance conforms to the following standards:
DIN EN 60601-1: 3/96 «Medical electrical equipment» –
Part 1: General requirements for safety
DIN EN 1060-1: 12/95, AMD 1 09/02 «Non-invasive sphygmomanometers» –
Part 1: General requirements
DIN EN 1060-3: 9/97 «Non-invasive sphygmomanometers» –
Part 3: Supplementary requirements for electro-mechanical blood
pressure measuring systems
This product conforms to the provisions of the EC directive 93 /42 / EEC
(Medical Device Directive).
0297