12
Storage and cleaning
Keep the unit in the storage case when not in use. Clean the product with a soft, slightly
damp cloth. Do not wet the cu or attempt to clean the cu with water. Do not wash
or dry clean the cu. Do not store the product in a place exposed to direct sunlight or
high humidity. Remove the batteries if the unit will not be used for 3 months or longer.
Always replace all of the batteries with new ones at the same time. If needed, please use
ultraviolet radiation for if disinfection.
Specifications
Model: BPW4100
Operation principle: Oscillometric method
Display: Liquid Crystal Display
Measuring range: 0-280mmHg (cu pressure)
Blood pressure
measurement: Systolic pressure: 60-280mmHg
Diastolic pressure: 30-250mmHg
Pulse rate: 40-180/min
Laboratory accuracy: Cu pressure: ±3mmHg
Pulse rate: ±5%
Ination: Automatic, soft adapted to blood pressure
Power supply: 2 batteries, type LR03(AAA)
Electric Shock Protection: Internal power supply appliance type BF
Operating Environment: Temperature +10 C / +40 C
Humidity 15 ~ 85%RH (Noncondensing)
Atmospheric Pressure 105 kPa ~ 80 kPa
Transport and Temperature -20C / +60C
Storage Environment Humidity 15 ~ 85%RH (Noncondensing)
Atmospheric Pressure 105 kPa ~ 80 kPa
Cu: Suitable for wrist circumference of 13.5 to 21.5cm
Accessory: Batteries(2*AAA)
If device is not used within specied temperature and humidity ranges the technical
accuracy can not be ensured.