4.2.4 Ramp Pressure
The ramp pressure setting is used for defining the start and end pressure when Ramp
Up and/or Ramp Down is configured.
IPAP shall start at 2 cmH
2
O above the configured Ramp Pressure. If the Ramp Pressure
is set below EPAP, the EPAP pressure will start at the configured Ramp Pressure.
CPAP mode:
Unit Min
Max Default Resolution Tolerance
cmH
2
O 2 EPAP or
CPAP
2 0.5 < 10
1.0 10
± (0.5 cmH
2
0 + 5%)
4.2.5 Humidifier
This setting defines whether the attachable humidifier shall be used.
WARNING!
Read the section 2.6 Humidification and Heating Warnings and Precautions, page 24
before activating the humidifier, to make sure all conditions are met and considered.
Min Max Default
Off
On
Off
4.2.6 Humidifier Level
This setting defines the level of humidification, if Humidifier is set to On.
Min Max Default
1 5 1
4.2.7 Heated Circuit
This setting defines whether a heated patient circuit shall be used as patient circuit.
WARNING!
Read the section 2.6 Humidification and Heating Warnings and Precautions, page 24
before activating the heated circuit, to make sure all conditions are met and
considered.
Treatment Functions and Settings
57
Doc. 007229 D-3 User manual