16 Safety Information
Vivo 50 Operating manual
Doc. 004970 En-Us C-2b
2.6 Humidification
Humidification must only be used if this has been prescribed by a
physician.
When using an external heated humidifier, it should be located below
the Vivo 50 and the patient to prevent injury from accidental spillage.
During transportation of the Vivo 50, the humidifier must be discon-
nected.
If the condensation in the patient circuit is excessive, the use of a
heated humidifier may require the installation of a water trap in the
circuit. The water trap prevents any condensated water in the patient
circuit from running into the patient airways and causing personal
injury.
When adding or removing a HME (Heat and Moisture Exchanger,
artificial nose) or HCH ((Hygroscopic Condenser Humidifier),
always perform a pre-use test.
Any humidifier connected to the Vivo 50 must comply with
ISO 8185.
Any HME connected to the Vivo 50 must comply with ISO 9360.
The use of an HME or an external humidifier may require readjust-
ment of the Vivo 50 low-pressure alarm.
Certain HME’s and HCH’s are sufficient to provide humidification
when the Vivo 50 is used invasively. Check specific suppliers’ recom-
mended use.
The Vivo 50 has been tested and validated with the Fisher & Paykel
MR850 heated humidifier.