EasyManua.ls Logo

Breg VPULSE - VPULSE Troubleshooting

Breg VPULSE
112 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
15
of 22
Troubleshooting
Frequently asked questions in
troubleshooting the VPULSE:
1. What should I do if the VPULSE does not
seem to get cold?
• Check the inside and outside of the water
reservoir for cracks, punctures or other signs
of damage. If damage is noted, please contact
Customer Care at Breg, 1.800.321.0607 or
1.760.795.5440.
• Make sure that the water container is properly
filled with water.
DO NOT operate the system with a water
container not filled with water or
not filled to the MINIMUM fill line, before
ice or Ice Bottles are added.
DO NOT operate the system with a water
container filled with water above the
MAXIMUM fill line, or with a tubing set
not properly connected.
• Check that enough ice is added to the water
container and that the tubing is not kinked.
• Check the power connection.
2. What should I do if the tubing set does not
connect to the VPULSE or pad?
• Check the tubing set and connectors for cracks,
punctures or other signs of damage. If damage
is noted, contact Customer Care at Breg,
1.800.321.0607 or 1.760.795.5440.
• Make sure the correct connectors are mating.
Check the color and number of connections.
• Make sure no foreign objects are blocking the
fluid connectors.
3. What should I do if the system does not turn
on?
Check wall adapter power connection at the
wall.
• Check the outlet to ensure it is properly
powered.
• Check wall adapter is connected at back of the
Control Unit.
• Make sure no foreign objects are blocking the
fluid connectors.
• Check that the front panel keys illuminate
when pressed.
• Check that the Start/Stop key ( ) illuminates
GREEN when pressed.
4. What should I do if the system is not pumping
water through the pad?
• Check power connection.
• Check that the tubing set is properly connected.
• Check that the pads are properly connected.
• Check that the water reservoir is properly filled
with cold water.
• Check that the Cold Therapy button has been
selected and the function illuminates WHITE.
• Check that the Start/Stop key ( ) illuminates
GREEN when pressed.
• Check that the tubing is not kinked.
• Start the system to ensure that the water is
flowing into the water reservoir. A return
stream is visible inside water reservoir.
5. What should I do if the system is not pumping
any air through the pad?
• Check power connection.
• Check that the tubing set is properly connected.
• Check that the pads are properly connected.
• Check that the front panel keys illuminate
when pressed and that a compression therapy
is selected. If the compression keys are both
continuously illuminated RED and VIOLET
the air pump might not be providing sufficient
pressure.
• If compression keys are both continuously
illuminated RED, the DATA card may not be
installed, the card may be full, or experiencing
a problem. Open the DATA door and reinstall
the DATA card.
• Check that the Start/Stop key ( ) illuminates
GREEN when pressed.
• Check that pad is inflating.
• Check the dynamic or sequential therapy mode
is enabled.
Refer to Breg Customer Care, 1.800.321.0607 or 1.760.795.5440 if
these actions are ineffective.

Other manuals for Breg VPULSE

Related product manuals