A brief summary of the tests carried out in accordance with EN60601-1-2:2007 is shown below.
The VPULSE is suitable for use in the specified electromagnetic environment. The consumer and/or user of the VPULSE
should ensure that it is used in an electromagnetic environment as described below:
The VPULSE uses RF energy only for its internal function. Therefore,
its RF emissions are very low and are not likely to cause interference
in nearby electronic equipment.
The VPULSE is suitable for use in all establishments, including
domestic establishments and those directly connected to the public
low-voltage power supply network that supplies buildings used for
domestic purposes.
Emissions Test Compliance Electromagnetic Environment Guidance
Group 1
Class B
Class A
Complies
RF Emissions
CISPR 11
RF Emissions
CISPR 11
Harmonic Emissions
IEC 61000-3-2
Voltage fluctuations / flicker emissions
IEC 61000-3-3
Portable and mobile RF communications equipment should be
used no closer to any part of the VPULSE, including cables, than the
recommended separation distance calculated from the equation
appropriate for the frequency of the transmitter.
Recommended Separation Distance
d=1,2√P
d=1,2√P 80 MHz to 800 MHz
d=2,3√P 800MHz to 2,5 GHz
where P is the maximum output power rating of the transmitter in
watts (W) according to the transmitter manufacturer and is the
recommended separation distance in metres (m).
Field strengths from fixed RF transmitters, as determined by an elec-
tromagnetic site survey (a), should be less than the compliance
level in each frequency range (b).
Interference may occur in the vicinity of equipment marked with the
following symbol:
Immunity Test Compliance Level Electromagnetic Environment Guidance
3 Vrms
3 V/m
80 MHz to
1 GHz
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
IEC 60601-1-2
Test Level
3 Vrms
50 kHz to
80 MHz
3 V/m
80 MHz to
2,5 GHz
Floors should be wood, concrete, or ceramic tile. If floors are covered
with synthetic material, the relative humidity should be at least 30%.
Main power quality should be that of a typical commercial and/or
hospital environment.
Main power quality should be that of a typical commercial and/or
hospital environment.
Immunity Test Compliance Level Electromagnetic Environment Guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
IEC 60601-1-2
Test Level
6 kV contact
8 kV air
2 kV for power
supply lines
1 kV for input/
output lines
1 kV differential
mode
2 kV common
mode
The VPULSE is suitable for use in the specified electromagnetic environment. The user of the VPULSE should ensure that it
is used in an electromagnetic environment as described below:
6 kV contact
8 kV air
2 kV for power
supply lines
N/A
1 kV differential
mode
N/A