Why are my BRONKHORST FLOW-BUS Measuring Instruments responding slowly to setpoint changes?
- SSandra AlexanderAug 15, 2025
If your BRONKHORST Measuring Instruments show no or slow response to setpoint changes, several factors could be at play. First, the slope value might be too high, affecting the ramp function for the setpoint signal; check and adjust it using a readout module or PC software. Secondly, the control mode or setpoint source may be incorrectly configured, directing the instrument to an unintended source; ensure the correct control mode and setpoint source are selected. A different instrument or PC on the FLOW-BUS could also be altering the setpoint; verify who has control over setpoint changes. Finally, incorrect polynomial factors could be the issue; confirm that the correct fluid is selected and compare the polynomial factors A through D against your calibration certificate.