VeraTemp
16
5–8 cm (2–3 inches)
away from forehead
Common Errors
If the screen displays
while on BODY mode, you may be
pressing the SCAN button before
aiming at the patient’s forehead.
If multiple attempts display the same results,
you may not be waiting the 5 seconds between uses to
allow the previous temperature to be completed.
To check previous temperatures, the memory
function (MEM) records the previous 32 scans with the
temperature reading and the mode in which it was taken.
If the temperature appears low, check the MODE.
If attempting a BODY temperature in the SURFACE Temperature
mode, Temperature may read 1.4-6.5°F lower than actual.