EasyManua.ls Logo

Brooklands VeraTemp - Temperature Variation and Accuracy Standards; Individual Temperature Variability; ASTM Accuracy Requirements

Brooklands VeraTemp
20 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6
Because of heat loss from various parts of the body that
are not covered by clothing, the normal temperature
measured on a persons forehead is usually lower than
covered parts of the body. The
American Society for Testing and Materials (ASTM) accuracy requirements
in the display range of 98°F–102°F (37°C–39°C) for infrared thermometers
is +/- 0.4 °F (+/-0.2 °C), whereas for mercury-in-glass and electronic
thermometers, the requirement per ASTM Standards E667–86 and
E1112–86 is +/- 0.2°F (+/-0.1 °C).
This infrared thermometer meets
Brooklands Incorporated
requirements established in ASTM
Standard (E1965-98). Full responsibility for the conformance of this product
temperature of the study group. Body temperature is
individual to the person and can vary on a healthy person
by more than +/- 1°C from 37°C (98.6°F). Your body
temperature is inuenced by many factors including but
not limited to the time of day, gender, age, ambient
temperature and metabolic activity.
VeraTemp™
Non-Contact Thermometers internal software
automatically adjusts the temperature to provide a
reading which approximates the body temperature.
Personal