EasyManua.ls Logo

BSV BS POOL PRO100 - Page 17

Default Icon
49 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
66 / 150
Manual PRO50-150 V.2
Additional recommendations:
- If possible, install a hose valve at input, and another one at output to ease the
cleaning and maintenance tasks.
-The probe holder water inlet can be connected after the filter, but then a good
maintenance of the filter cleaning must be ensured, otherwise the measure can
be affected by the chlorine consumption inside the filter.
-As an alternative, the inlet can be connecter before the filter, but then it is
strictly necessary to install a specific cartridge filter to avoid the probe
contamination.
3.2.5.3.2 Probe set up
Before installing the probe, it is necessary to fill the cap with electrolyte. Please,
follow carefully the following steps:
1)
Unscrew the header cap from the sensor body and keep both parts in a
clean and stable surface.
Important: Don’t touch the membrane and the gold electrode, they could
be contaminated.
2)
Fill the header cap with the supplied electrolyte. Fill it gently to avoid
creating bubbles.

Related product manuals