Digestion Unit K-424
5 Operation
5.4 Carrying out the Kjeldahl digestion
Switch the unit on and preheat for approx. 5 minutes
on stage 10.
Wear protective goggles!
Place the digestion tubes in the digestion tube holder
on the rack .
Introduce the sample and reagents into the digestion
tubes .
Seal off unused connecting branches with the glass
cap (item No. 17396) .
Never install empty digestion tubes!
Place the suction module with the fitted gaskets on
the digestion tube holder.
Turn on scrubber or water jet pump bevor any gases
can develop.
5.5 Malfunctions
Malfunction Possible cause Remedy
Heating elements do not 1. Mains plug not plugged in - Connect unit to the mains
heat up 2. Unit switched OFF - Switch ON
3. Energy regulator in position „0“ - Set to position "10"
4. Energy regulator faulty - Replace regulator
5. Heating elements faulty - Replace heating element
Lamp does not light up (H 1) Lamp faulty Replace lamp
Table 4: Malfunctions
Fig. 10: Operation of the K-424