Bualo Filter ViroVac™ Operator’s Manual: Page 12
2.2 ViroSafe® Filter Instructions
Note: Before installing or removing any lter, be sure that the system is turned o.
Filter Installation Instructions:
e installation of the ViroSafe® Filter into the BUFFALO FILTER ViroVac™ Smoke Evacuation
System(s) is quick and simple.
1. Remove the ViroSafe® Filter from the shipping box and discard any protective wrapping. Ex-
amine all lters for damage during shipping and storage. Do not install any lter with visible
signs of structural damage.
2. Insert the ViroSafe® Filter into the lter receptacle. Be sure that the lter is seated completely
against the bottom of the lter chamber and clip is fully engaged.
WARNING: is device is not intended for evacuation of uid. If uid is expected to be as-
pirated using the ViroSafe® Filter or the BUFFALO FILTER ViroVac™ system, uid collection
devices must be installed with the vacuum hose assembly. Failure to install a uid collection
device may cause lter blockage and/or electrical damage.
Filter Removal Instructions:
1. Aer the ViroSafe® Filter has been exhausted and requires changing, turn the smoke
evacuation system o and disconnect any accessory tubing attached to the lter.
2. Depress the lter tab and pull the ViroSafe® Filter from the smoke evacuation system and dis-
pose of in accordance with hospital policy. e ViroSafe® Filter may be disposed of or inciner-
ated.
3. Clean the unit with appropriate germicide prior to re-use. Follow the indicated instructions
for maintenance and installation of a new ViroSafe® Filter.
CAUTION: Using any other lter or accessory not supplied by BUFFALO FILTER may cause
damage to the system and/or cause the system to be inoperable and may void the warranty.
WARNING: e ViroSafe® Filter should be changed when the Filter Life Indicator shows red
ashing LED (replace). Failure to change this lter may result in decreased eciency and
contamination of the electric motor, vacuum pump, and sound absorbing media within the
system, or non-operation of smoke evacuator.
Bualo Filter P/N: VV120 & VV220
ViroSafe® Filter P/N: VS353
Conguration: Portable or tabletop
ViroSafe® Filters - Multi Port Filter: 4-Stage Filtration In One Casing,
(Pre-Filter, ULPA, Carbon, Post-Filter)
Filter(s): ULPA
Particle Size, µm: 0.1 to 0.2 Microns At 99.999% Eciency
Filter Life: Automatic Factory Set Filter Sensor
Filter Life Indicator: Timed Replacement

Related product manuals