BW Clip
2 OR 3 YEAR H
2
S, CO, O
2
, SO
2
SINGLE-GAS DETECTOR
Part Number 50115903-067 | OM-EN-PALO-A4 | BW Clip Operator Manual
BW Clip is a portable, single-gas detector manufactured by
BW Technologies by Honeywell (BW). BW Clip continuously measures
the concentration of a specic gas in the ambient environment, and
activates an alarm when concentrations exceed alarm setpoints. It is your
responsibility to respond appropriately to the alarms.
BW Clip is a member of the GasAlert family of products. It is compatible
with IntelliDoX and MicroDock II automatic test and calibration stations,
as well as Fleet Manager II software version 4.0 or higher. For more
information, visit the product website at www.honeywellanalytics.com
Safety Information: Read First
1. Substitution of components may impair Intrinsic Safety.
2. Periodically test the response of the sensor by exposing the detector
to a target gas concentration that exceeds the low alarm setpoint.
Manually verify that the audible and visual alarms are activated.
3. Detectors in hibernation mode are not functional.
1. Activate the detector before the activation date on the package.
2. This product is a gas detector, not a measurement device.
3. Ensure that the sensor grill is free of dirt, debris, and is not obstructed.
4. Clean the exterior with a soft, damp cloth.
5. Bump test the detector in a normal atmosphere that is free of
hazardous gas.
6. For optimal performance, periodically zero the sensor in a normal
atmosphere (20.9% v/v O
2
) that is free of hazardous gas.
Informations de sécurité – À lire au préalable
AVERTISSEMENT
1. Le remplacement d’un composant de l’appareil peut compromettre la
sécurité intrinsèque du détecteur.
2. Testez périodiquement la réponse du capteur en exposant le détecteur
à une concentration de gaz supérieure au seuil d’alarme basse. Vériez
manuellement que les alarmes sonores et visuelles sont activées.
3. Détecteurs en mode veille prolongée ne sont pas fonctionnels.
MISE EN GARDE
1. Activez le détecteur avant la date d’activation sur l’emballage.
2. Ce produit est un détecteur de gaz, pas un dispositif de mesure.
3. Veillez à assurer que la grille du capteur est exempt de saleté, les
débris, et ne soit pas obstrué.
4. Nettoyez l’extérieur de l’appareil à l’aide d’un chiffon doux et humide.
5. Veillez à effectuer un test fonctionnel du capteur dans une atmosphère
normale qui est exempte de gaz dangereux
6. Pour des performances optimales, zéro périodiquement le capteur
dans une atmosphère normale (20,9% v/v O
2
) qui est exempte de gaz
dangereux.
BW Clip Factory Calibration Certicate
This instrument was factory inspected, tested and calibrated in accordance with the conditions and
requirements of our registered Quality System, Operating Standards and Sales Agreements.
Getting Started
PushbuttonPushbutton
Sensor and
sensor grill
Sensor and
sensor grill
Bubble label
with product name
and gas type
Bubble label
with product name
and gas type
Audible
alarm
Audible
alarm
LCDLCD
LED
visual alarms
LED
visual alarms
Infrared port
Alligator
clip
Product label
LCD and Symbols
Pushbutton prompt. When this symbol is displayed,
press and hold the pushbutton until the symbol
disappears.
BW Clip O
2
models only. Oxygen concentration is
measured as percent by volume.
BW Clip H
2
S, CO and SO
2
models only. Toxic gas
concentration is measured in parts per million.
Low alarm and high alarm. These symbols are
displayed when a gas concentration that exceeds
alarm setpoints is detected.
Non-compliance warning. This symbol is displayed
when a diagnostic or compliance test fails; when the
operating life countdown reaches 24 hours or less;
and for up to 30 days after operating life expiry. It is
also displayed when certain notications are enabled
through Fleet Manager II via an IntelliDoX station.
Real-time clock. This symbol is displayed when real-
time clock is enabled through Fleet Manager II via an
IntelliDoX station.
End of operating life warning. This symbol is displayed
when the operating life countdown reaches 24 hours
or less.
Peak reading. This symbol is displayed when a
maximum gas exposure event has occurred within the
past 24 hours.
Bump test reminder. If the detector is congured to
display a bump test reminder, this symbol is displayed
when a bump test is due. indicates a failed test.
BW Clip O
2
models only. This symbol is displayed
when more than 24 hours have elapsed since the last
successful zero procedure was performed. indicates
a failed test.
Operating life remaining in months (MM) or days (DD).
Question and Answer IconNeed help?

Do you have a question about the BW Technologies BW Clip and is the answer not in the manual?

BW Technologies BW Clip Specifications

Operating Environment IconOperating Environment
operating temperature H2S-40° to +50°C (-40° to +122°F)
operating temperature CO-30° to +50°C (-22° to +122°F)
operating temperature O2-20° to +50°C (-4° to +122°F)
operating temperature SO2-30° to +50°C (-22° to +122°F)
operating humidity5% to 95% relative humidity (non-condensing)
Detection and Alarms IconDetection and Alarms
detection range H2S0 to 100 ppm
detection range CO0 to 300 ppm
detection range O20 to 25% by volume
detection range SO20 to 100 ppm
audible alarm≈ 95 dB at 10 cm (3.9 in)
alarm setpoints H2S low10 ppm
alarm setpoints H2S high15 ppm
alarm setpoints CO low35 ppm
alarm setpoints CO high200 ppm
alarm setpoints O2 low19.5% vol
alarm setpoints O2 high23.5% vol
alarm setpoints SO2 low5 ppm
alarm setpoints SO2 high10 ppm
Battery and Power IconBattery and Power
battery typeLithium, non-replaceable
Sensor and Display IconSensor and Display
sensor typeElectrochemical cells
display typeLiquid crystal display (LCD)
Weight and Dimensions IconWeight and Dimensions
instrument weight92 grams (3.2 ounces)
instrument dimensions41 x 50 x 87 millimeters (1.6 x 2.0 x 3.4 inches)

Summary

Getting Started

Device Components Overview

Identifies key physical components of the BW Clip detector like LCD, alarms, pushbutton, and sensor.

Understanding LCD Symbols and Indicators

Explains the meaning of various symbols displayed on the detector's LCD for different operational states and warnings.

Detector Operation Modes

Activating a New Detector

Step-by-step guide for the initial activation and internal diagnostic testing of a new BW Clip detector.

Normal Operating Mode

Describes the detector's state when functioning normally, displaying gas type and remaining operating life.

Operating Life Countdown

Details how the detector displays remaining operating life in months, then days, and finally hours before expiry.

End of Operating Life

Explains the detector's state when its service life has ended, disabling safety functions.

Internal Diagnostic Tests and Fail-Safe

Covers automatic internal tests, the diagnostic fail-safe sequence, and resulting detector deactivation.

Alarms and Warnings

Gas Exposure Alarms

Details low, high, and over/under-limit gas concentration alarms, including audible and visual cues.

Specific Model Reminders

Explains the automatic zero reminder for O₂ models and peak reading display.

Non-compliance Indicators

Describes the non-compliance symbol and LED patterns for diagnostic failures, failed tests, or expiry warnings.

Advanced Features and Settings

Recalling Events and Settings

How to cycle through stored events like peak readings, alarms, and settings using the pushbutton.

Zeroing the Sensor

Procedure for calibrating the sensor baseline, especially important for O₂ models.

Bump Testing Procedure

Information on performing bump tests to verify sensor response and functionality.

Event Logging

Details the types of events logged by the detector and how to transfer them.

Optional Detector Settings

Enabling features like displaying gas readings during alarms or real-time clock via IntelliDoX.

Hibernation Mode

Explains how to suspend a detector's operation for extended periods using a hibernation case or software.

Product Specifications and Compliance

Publication Information

Details about the operator guide, its scope, and copyright information.

Operating Life and Detection Ranges

Specifies detector service life based on usage and the measurement ranges for different gases.

Factory Alarm Setpoints

Lists the default low and high alarm concentration setpoints for H₂S, CO, O₂, and SO₂ detectors.

Ratings and Certifications

Details the safety standards and certifications the BW Clip detector complies with.

Technical Specifications

Provides details on shelf life, weight, dimensions, operating conditions, and alarm types.

Warranty and Liability

Outlines the product's limited warranty, exclusions, and limitations of liability.