The table below summarizes the material groups and their attributes.
* means reference material for that
Low power, heater and guard 90 mA each.
High power, heater 105 mA and guard 125 mA.
All groups use start time 0.3s and stop time 0.8s.
The TCi sensor has a solid and hard surface. When measuring solids, there will always be some
tact resistance, which may significantly affect the results. The quality of heat transfer depends
on many parameters such as type of material, surface quality, etc.
To overcome this problem we use a contact agent (CA). The best CA available is water, sinc
has relatively high thermal conductivity (~0.6 W/mK), low viscosity, and is easy to apply and
clean. Water can be used in a limited temperature range though, from ~5
Calibrations of solids (except foams) are all done with water.
C a different CA is needed. Any such
material will have a lower thermal conductivity than water. And since calibrations are done only at
room temperature with water (and automatically corrected at other temperatures
to correct for the contact resistance. This correction factor (CF) may be constant over the entire
range of materials, or may have a simple relationship with regard to the measured