TECHNICAL CHARACTERISTICS
TYPOLOGY (MDD 93/42/EEC) Class IIa Medical Device
MODEL
MIKO
POWER FEEDING 230V~ / 50Hz
POWER CONSUMPTION 170VA
FUSE F 1 x 1.6A 250V
MAX PRESSURE 220 kPa (2.2 Bar)
MAX AIR FLOW 13 l/min
OPERATING PRESSURE 95 kPa (0.98 Bar)
OPERATING AIR FLOW 4.8 l/min a 95kPa
NEB-RATE (with 4ml of 0.9% NaCl solution) 0.35 ml/min
MMAD 2.44 µm
GSD 2.87
WEIGHT 1.45 Kg
SIZE 148 x 171 x 114 (h)
NOISE LEVEL
(measured as specifications of EN 3544-1)
Approx. 56dB (A)
DUTY CYCLE
(to 40°C and 110% operating voltage)
NOT-STOP OPERATED
MIN. CAPACITY NEBULIZER 2ml
MAX. CAPACITY NEBULIZER 6ml
WORKING CONDITION
Room temperature: 10 ÷ 40°C
Room humidity percentage: 20 ÷ 85% RH
Altitude: 0 ÷ 2000m s.l.m.
CONSERVATION CONDITION AND TRASPORT
Room temperature: -25÷ 70°C
Room humidity percentage: 10 ÷ 95% RH
MMAD = Mass Median Aerodynamic Diameter
GSD = Geometric Standard Deviation
CUT OFF
( um)
5.2 l/min
4.0 l/min