CAIRE FreeStyle - User Manual

CAIRE FreeStyle
91 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Loading...
!
!
!
!
!
!

Other manuals for CAIRE FreeStyle

Questions and Answers

  • R
    Richard CookJul 27, 2025
    What to do if CAIRE Oxygen Equipment FreeStyle has a continuous alarm and red light?
    • B
      brian10Jul 27, 2025
      If your CAIRE FreeStyle Oxygen Equipment sounds a continuous alarm and the red alarm light is on, it means the device hasn't detected a breath for 15 minutes. Check the cannula connection, ensure the tubing isn't kinked, and make sure the cannula is correctly positioned so you are breathing through your nose. If you breathe through your mouth, your clinician might suggest using a chin strap.
  • C
    cory36Jul 31, 2025
    What to do if CAIRE Oxygen Equipment FreeStyle has an intermittent alarm and yellow light?
    • D
      Damon MaddoxAug 1, 2025
      If your CAIRE FreeStyle Oxygen Equipment has an intermittent alarm and the yellow BATT light is flashing, it means the battery needs charging. Connect the device to a DC or AC outlet within 2 minutes, or replace the battery with a fully charged one.
  • S
    savannah50Aug 5, 2025
    How to fix CAIRE FreeStyle Oxygen Equipment that does not operate?
    • K
      Kelly GreeneAug 6, 2025
      If your CAIRE FreeStyle Oxygen Equipment doesn't turn on when you press a pulse flow selection button, the battery is likely discharged. Try powering the unit through a DC or AC outlet. Contact your equipment provider and switch to an alternate oxygen source if needed.
  • J
    James EsparzaAug 11, 2025
    What to do if the yellow BATT light is flashing intermittently on my CAIRE Medical Equipment?
    • K
      Kim EdwardsAug 11, 2025
      If your CAIRE Medical Equipment's battery requires charging and triggers an intermittent alarm with the yellow BATT light flashing, connect it to a DC or AC outlet within 2 minutes. Alternatively, replace the current battery with a fully charged one.
  • R
    Ronald ClarkAug 16, 2025
    What to do if CAIRE Medical Equipment has a continuous alarm and the red light is on?
    • J
      James WilsonAug 16, 2025
      If your CAIRE Medical Equipment alarm sounds continuously and the red alarm light stays on, it means the device hasn't detected a breath for 15 minutes. Check the cannula connection to ensure it's properly connected and that the tubing isn't kinked. Also, make sure the cannula is correctly positioned, and you are breathing through your nose. If you breathe through your mouth, your clinician might suggest using a chin strap.
  • J
    James ThompsonAug 21, 2025
    What to do if my CAIRE Medical Equipment does not operate when pressing the pulse flow selection button?
    • R
      Robert MackAug 21, 2025
      If your CAIRE Medical Equipment does not operate when you press a pulse flow selection button, it could be due to a discharged battery or a malfunction. Try powering the unit through a DC or AC outlet. If the problem persists, contact your Equipment Provider and switch to another source of oxygen if necessary.

Summary

FREESTYLE AND FREESTYLE 5 SPECIFICATIONS

FREESTYLE AND FREESTYLE 5 ALARMS

FREESTYLE AND FREESTYLE 5 CLASSIFICATIONS

HOW THE FREESTYLE AND FREESTYLE 5 WORK

FREESTYLE AND FREESTYLE 5 FLOW RATES

INDICATIONS FOR USE

CONTRAINDICATIONS

FREESTYLE AND FREESTYLE 5 OPERATION CHECK

OXYGEN CONCENTRATION TEST AND SPECIFICATION

SAFETY GUIDELINES AND OPERATIONAL SAFETY, WARNINGS, AND CAUTIONS

CONNECTING TO EXTERNAL POWER

SHOWING PANEL BUTTONS, INDICATORS, AND ALARMS

CLEANING THE AIR INTAKE; GROSS PARTICLE FILTER

CLEANING THE CABINET AND POWER SUPPLIES

EQUIPMENT PROVIDER MAINTENANCE

SERVICE

Tools for Freestyle and Freestyle 5 Service and Maintenance

Lists and illustrates essential tools required for servicing and maintaining the concentrators.

COMPONENTS

PSA Removal

Step-by-step instructions for safely removing the Pressure Swing Adsorption (PSA) system.

Compressor Replacement

Procedures for replacing the compressor unit within the concentrator.

Product Manifold Replacement

Steps to remove and replace the product manifold assembly of the concentrator.

Sieve Bed Replacement

Procedures for removing and replacing the sieve bed assemblies in the concentrator.

MEASURING THE SYSTEM PRESSURE

PRESSURE PROFILES

FREESTYLE HOUR METER

MOTOR CONTROLLER BOARD REPLACEMENT

CIRCUIT BOARD REPLACEMENT

CONTROL PANEL; KEY PAD REPLACEMENT

FAN REPLACEMENT

TROUBLESHOOTING

General Troubleshooting

Initial steps and checks to perform before consulting a troubleshooting chart for malfunctions.

CAIRE FreeStyle Specifications

General IconGeneral
AC Power100-240 VAC, 50/60 Hz
Audible AlarmYes
Battery Recharge Time4 hours
Power SupplyAC/DC adapter, rechargeable battery
Operating Temperature5°C to 40°C (41°F to 104°F)
Storage Temperature-4° to 140°F (-20° to 60°C)
Dimensions8.6 in H x 6.1 in W x 3.5 in D (21.8 cm H x 15.5 cm W x 8.9 cm D)
Oxygen Concentration87% to 96%
Humidity Range15 to 95% non-condensing