Programming and Menu
Selection Buttons
Status LEDs
Red: Error
Yellow: Hold/Pause
Green: Normal Operation
Kangaroo
TM
ePump Enteral Feeding
Pump "Flushing Set Required"
Power Button
QUICK
REFERENCE
GUIDE
FLUSHING MODE
OPERATION
SYMPTOM PROBABLE CAUSE CORRECTIONS
Hold Error
Pump in Holding mode
for more than 10 minutes.
Press Continue to return to Holding screen, or press and hold the
Power button to turn off. Several options are available from the Hold-
ing screen, such as Run and Adjust Settings.
Flow Error
Occlusion in pump-patient
line (downstream occlusion).
Check for and correct the occlusion in the line between pump and
patient. Replace Pump Set if error can’t be resolved.
Feed Error
Empty bag or occlusion
in bag-pump line
(upstream occlusion).
Check for empty bag and refill, or check for and correct the occlusion
in the line between the pump and feed bag. Replace Pump Set if error
can’t be resolved.
Flush Error
Empty bag or occlusion in
flush line.
Check for empty bag and refill, or check for and correct the occlusion
in the line between the pump and flush bag. Replace Pump Set if error
can’t be resolved. Check the valve pocket in the pump set loading area
for dampness or dirt.
Pump Set
Use >24 Hrs
Pump Set in use beyond
recommended length of time.
Informational warning message blinks on upper left of screen. Its
recommended that the Pump Set be replaced.
Pump Set
Dislodged
Pump Set not properly
loaded to pump.
Check black ring retainer (MISTIC) and valve; reload Pump Set and
restart pump to continue. Replace Pump Set if error can’t
be resolved.
Battery Low
Battery charge too low.
Immediately plug pump into A/C outlet to recharge battery and to
provide power for pump.
Rotor Error
Pump Set tubing not properly
loaded on rotor, or other unusual
rotor operating condition.
Check that the tubing is loaded on the rotor and that it’s not dam-
aged, torn, etc. Reload tubing or replace Pump Set.
System Error
General error caused by many
factors. See manual for error list.
Power down pump and attempt restart to clear error. Call customer
service at 800.962.9888 with screen code if error can’t be resolved.
For additional operating instructions and information, see the operating manual for the Kangaroo™ ePump Enteral Feeding Pump,
Flushing Mode Operation "Flushing Set Required."
Below are some of the probable causes of alarm conditions and how to correct them.
© 2018 Cardinal Health. All Rights Reserved. CARDINAL HEALTH, the
Cardinal Health LOGO are trademarks of Cardinal Health and may be
registered in the US and/or in other countries. 01/2018 - H9817 - WF#2098413
15 Hampshire Street 800.962.9888
Mansfield, MA 02048 508.261.8000

Other manuals for Cardinal Health Kangaroo ePump

Question and Answer IconNeed help?

Do you have a question about the Cardinal Health Kangaroo ePump and is the answer not in the manual?

Cardinal Health Kangaroo ePump Specifications

General IconGeneral
BrandCardinal Health
ModelKangaroo ePump
CategoryMedical Equipment
LanguageEnglish

Related product manuals