6. Close blue door. Display
should read SETLOADED.
3. With pump display facing you
open blue door on right hand side
of pump.
4. Grasp finger tab on valve and insert
valve into valve slot on the left.
5. Grasp black ring retainer, and gently
wrap tubing anti-clockwise around
rotor. Pull up retainer and insert
into slot.
1. Locate power button on lower
right hand corner of control panel.
Press to turn pump on.
2. Confirm if you want to KEEP or
CLEAR any prior pump settings.
Pump Set Loading Procedure.
Kangaroo™ ePump Enteral Feeding Pump
User Guide (Feed Only Mode)

Other manuals for Cardinal Health Kangaroo ePump

Question and Answer IconNeed help?

Do you have a question about the Cardinal Health Kangaroo ePump and is the answer not in the manual?

Summary

Pump Set Loading Procedure

Powering On the Pump

Locate and press the power button on the lower right corner of the control panel to turn the pump on.

Confirming Prior Settings

Confirm whether to KEEP or CLEAR any previously saved pump settings.

Opening the Pump Door

With the pump display facing you, open the blue door on the right-hand side of the pump.

Inserting the Valve

Grasp the finger tab on the valve and insert it into the valve slot on the left side.

Loading the Tubing Rotor

Wrap tubing anti-clockwise around the rotor, then insert the retainer into its slot.

Completing Set Loading

Close the blue door; the display should indicate 'SET LOADED'.

Set Priming and Pump Programming Procedure

Priming the Pump

Press PRIME PUMP, then AUTO PRIME or HOLD TO PRIME. Press DONE when complete.

Adjusting Feed Rate

Press ADJUST FEED, then FEED RATE to set the hourly delivery rate.

Programming Hourly Feed Rate

Use the left buttons to program the hourly feed rate and press ENTER to confirm.

Programming Feed Volume

Press FEED VTBD to program the Volume To Be Delivered.

Starting and Monitoring Feeding

Setting Final Feed Volume

Use the left buttons to program the feed volume and press ENTER to confirm.

Commencing Feeding Regime

Press DONE to return to the main screen, then press RUN to start feeding.

Feeding Container Height Requirement

The bottom of the feeding set container must be 46 cm above the pump for accuracy.

Alarm and Error Code Information

Hold Error

Pump left in Hold for over 10 minutes. Press CONTINUE or POWER DOWN.

Rotor Error

Issue with pump set tubing around the rotor. Ensure silicone tubing is correctly loaded.

Flow Error

Obstruction downstream. Check for kinks or blockages in tubing below the pump.

Feed Error

Container empty or upstream obstruction. Check tubing or attach new container.

Feeding Complete

Programmed feed has been delivered. Power down the pump.

Pump Set Dislodged

Retainer ring not loaded properly. Ensure pump set is correctly loaded.

Low Battery

Battery needs immediate recharging. Plug the pump into an AC outlet.

System Error

General error. Consult pump manual or send pump for service.