1780-M057-O1 Rev C Hydraulic Floor Stand Tank/Hopper
11
INDICATOR INSTALLATION
1. Install the digital indicator.
2. The indicator cable is terminated at the Surge Suppression box located in the front door of
the Totalizer enclosure. Make sure the surge suppression box is grounded back to the
indicator and the AC power ground. Failure to provide a good ground will make the surge
suppression inoperative. See Figure No. 11.
Figure No. 11 – Ground Lug Location
The Ground wire must be attached to the ground lug on the Suppressor
Package. It should be run with the indicator cable and terminated to the ground
lug on the weight indicator.
3. Power up the indicator, remove the signal lines from the summing card and read the signal
from each transducer. Any transducer/cell pair that has a low output should be inspected to
make sure it is taking load. If necessary place a small shim under the lower frame cell pad.
DO NOT ADD OIL TO THE CELL!!
Ground Lug