INSTALL
MEMO-CHIP
RUN TEST MEMORY UTILITY SETTINGS
This section is
concerned
with running
client blood
tests
This section is
used to
recall
previous
results
This section is
used to
conduct the
OPTICS
CHECK
This section is
used to
adjust time,
date & units
INSERT STRIP
APPLY
TESTING
CHEM
CONTROL
EXIT
CHECK STRIP
RUN
CLEAR
MEMORY
PRINT
SOUND
SET TIME
EXIT
SET DATE
UNITS
EXIT
Using the CardioChek
®
PA Analyser
The Navigation Menu
Use this selection to undertake an OPTICS CHECK
at the beginning of every day when you plan to
conduct tests.
Use this selection to CONDUCT CLIENT BLOOD TESTS. When you have
inserted a strip, the display will automatically change to APPLY SAMPLE.
Once the sample is applied, the display will change to TESTING.
Use this selection to RECALL client results. Useful if the analyser switches
off during the health check or you need to refer back to previous tests.
The previous 30 test results are stores by date and time.
RESULTS
2½ mins
Quick Guide