Safety Designation per IEC 60601-1
Type of protection against
electrical shock
Internally powered
Equipment
Degree of protection against
electrical shock
Type B applied part
Power Equipment Battery type: PP3 battery,
Alkaline Manganese Dioxide,
9.0V, 550mAh
Battery life: 2000 tests.
Degree of Electrical connection
between equipment and Patient
Equipment designed as non-
electrical connection to the
patient.
Degree of mobility Transportable
Mode of operation Continuous
Classifications according to IEC 60601-1
MicroCO meter Applied part, type B
WARNING: No modification of this equipment is allowed.
WARNING: Do not connect devices that are not specified as part of
the system.
NOTE: When you connect other equipment to the unit, always make
sure that the whole combination complies with the international safety
standard IEC 60601-1-1 for medical electrical systems. When
connecting to a PC with the supplied serial cable then the PC must be
IEC 60601-1 / ANSI/AAMI ES60601-1:2005 / CAN/CSA-C22.2 No.
60601-1:14 compliant.
WARNING: The user must not touch any voltage-carrying parts
and the patient at the same time
.