32
IEC61000-4-4
± 1 kV for input
/ output lines
Serial cable
<3m)
hospital environment.
Surge
IEC61000-4-5
± 1 kV line(s) to
line(s)
± 2 kV line(s) to
earth
Not Applicable
(battery
Powered)
Mains power quality
should be that of a
typical commercial or
hospital environment.
Voltage dips,
short
interruptions
and voltage
variations on
power supply
input lines
IEC61000-4-11
<5% U
T
(>95 % dip in
U
T)
For 0.5 cycle
40% U
T
(60 % dip in U
T)
for 5 cycles
70 % U
T
(30 % dip in U
T
)
for 25 cycles
<5% U
T
(>95 % dip in
U
T
)
For 5 s
Not Applicable
(battery
Powered)
Mains power quality
should be that of a
typical commercial or
hospital environment. If
the user of the MicroCO
requires continued
operation during power
mains interruptions, it is
recommended that the
MicroCO be powered
from an uninterruptable
power supply or a
battery.
Power
frequency
(50/60Hz)
Magnetic field
IEC61000-4-8
3 A/m 3 A/m If incorrect operation
occurs, it may be
necessary to position
the MicroCO further
from sources of power
frequency magnetic
fields or to install
magnetic shielding. The
power frequency
magnetic field should be
measured in the
intended installation
location to assure that it
is sufficiently low.
NOTE U
T
is the a.c. mains voltage prior to application of the test level.