Servicing
If your unit requires service or repair please see page 73 for contact
details. A full service manual including circuit diagram and parts list is
available upon request.
Trouble Shooting Information
Should you encounter problems operating your MicroCO meter, please
consult the table below:
Problem Possible Cause Solution
MicroCO cannot be
switched on or “bat”
is displayed
Batteries are flat Change the battery
Battery life is
shorter than
expected
Unit not being
switched off
Turn the unit off after
use
“CEL” is displayed Fuel cell is
depleted
Return unit for cell
replacement
“Err” is displayed Gas cylinder empty Check valve on cylinder
for contents and replace
cylinder if necessary
Check the calibration
value on the cylinder is
20ppm.
“bt2” is displayed Internal battery
has expired
Return unit for battery
replacement
“gAS” is displayed Fuel cell is
depleted
Return unit for cell
replacement
Fuel cell is
contaminated by
residual gas or
solvents
Allow unit to be
exposed to clean air.
Turn unit ON and await
message to disappear,
if not return unit for cell
replacement.