66
WARNING:
The MicroCO should not be used adjacent to or stacked with other
equipment. If adjacent or stacked use with other equipment is necessary, the
MicroCO and the other equipment should be observed /monitored, to verify
normal operation in the configuration in which it will be used.
Guidance and Manufacturer’s Declaration
Electromagnetic Emissions
The MicroCO is intended for use in the electromagnetic environment specified
below. The customer or the user of the MicroCO should assure that it is used in
such an environment.
Emission Test Compliance Electromagnetic Environment -
Guidance
RF emissions
CISPR 11
Group 1
The MicroCO uses RF energy only
for its internal function. Therefore,
its RF emissions are very low and
are not likely to cause any
interference in nearby electronic
equipment.
RF emissions
CISPR 11
Group B
The MicroCO is suitable for use in all
establishments, including domestic
establishments and those directly
connected to the public low-voltage
power supply network that supplies
buildings used for domestic
purposes
Harmonic emissions
IEC61000-3-2
Not Applicable
(battery
powered)
Voltage fluctuations /
flicker emissions
IEC61000-3-3
Not Applicable
(battery
powered)
Guidance and Manufacturer’s Declaration
Electromagnetic Imm
unity
The MicroCO is intended for use in the electromagnetic environment specified
below. The customer or the user of the MicroCO should assure that it is used in
such an environment.
Immunity
Test
IEC 60601 test
level
Compliance
Level
Electromagnetic
Environment -
Guidance
Electrostatic
discharge
(ESD)
IEC61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV
contact
± 8 kV air
Floors should be wood,
concrete or ceramic tile.
If floors are covered
with synthetic material,
the relative humidity
should be at least 30%.
Electrical fast
transient /
± 2 kV for
power supply
Not Applicable
(Battery
Mains power quality
should be that of a

Related product manuals