29
The MicroLoop uses a Digital Volume Transducer, an extremely
stable form of volume transducer, which measures expired air
directly at B.T.P.S (Body Temperature and Pressure with Saturated
water vapour) thus avoiding the inaccuracies of temperature
corrections. The transducer is insensitive to the effects of
condensation and temperature and avoids the need for individual
calibration prior to performing a test.
Test results may be uploaded to a PC using the optional Spirometry
PC Software
and patient details may be downloaded to the
MicroLoop.