Operation – SNIP (Sniff Nasal Inspiratory
Pressure)
Insert the battery into the rear of the MicroRPM.
Fit the Nasal Probe Adapter into the MicroRPM and then attach the
correct size Nasal Probe, as shown below.
The correct size (1-4) can be ascertained by fitting a Nasal Probe to
the unit, then firmly inserting the Nasal Probe into a nostril. Instruct
the subject to block the open nostril with a finger and then to
attempt a sniff. The correct Nasal Probe size has been selected once
there is no leakage around the Nasal Probe.
Nasal probe
2
Microcomputer unit
Nasal plug
adapter