What does error I 5 mean on my Carestream CS 8100SC Dental equipment?
- KKristy LeonardAug 1, 2025
The position of the head clamps do not match the selected exam program. Position the head clamps manually for the selected exam.
What does error I 5 mean on my Carestream CS 8100SC Dental equipment?
The position of the head clamps do not match the selected exam program. Position the head clamps manually for the selected exam.
What does error I 17 mean on my Carestream CS 8100SC?
The sensor is in the positioning process for either panoramic, 3D or Ceph mode. Wait for the end of the positioning process.
What does error I 9 mean on my Carestream CS 8100SC?
The Cephalometric sensor is in the cooling process. Wait until the Ready Indicator LED is green.
Why is my Carestream Dental equipment X-ray tube cooling?
The X-ray tube is cooling. Please wait until the Generator Cooling Indicator on the Acquisition interface reaches zero.
What to do if my Carestream CS 8100SC shows thermal security issues?
The system is undergoing a cooling process for thermal security. Wait until the Generator Cooling Indicator on the Acquisition interface reaches zero.
What to do if the handswitch was released too early on my Carestream CS 8100SC?
The user may have released the exposure button prematurely. Relaunch the acquisition and hold the exposure button until the end of the acquisition cycle.
When to release the handswitch on Carestream CS 8100SC Dental equipment?
The acquisition has ended. Release the exposure button of the X-ray remote control.
Specifies mandatory checks for computer system compatibility with CS 8100SC software.
Enables selection of different radiological exams and acquisition formats.
Allows selection of patient parameters affecting image quality based on age and morphology.
Provides step-by-step instructions for powering on the CS 8100SC unit and establishing connection.
Details preparation, parameter setting, and launching for lateral cephalometric image acquisition.
Details preparation, positioning, and launching for frontal AP/PA cephalometric image acquisition.
Details preparation, positioning, and launching for submento-vertex cephalometric image acquisition.
Details preparation, positioning, and launching for carpus image acquisition (optional).
Provides quick guidance for common malfunctions and error messages displayed on the acquisition interface.